Research Study Abstract

GT3X+ accelerometer placement affects the reliability of step-counts measured during running and pedal-revolution counts measured during bicycling

  • Published on Oct. 21, 2015

Background: Accelerometers provide a measure of step-count. Reliability and validity of step-count and pedal-revolution count measurements by the GT3X+ accelerometer, placed at different anatomical locations, is absent in the literature. The purpose of this study was to investigate the reliability and validity of step and pedal-revolution counts produced by the GT3X+ placed at different anatomical locations during running and bicycling.

Methods: Twenty-two healthy adults (14 men and 8 women) completed running and bicycling activity bouts (5 minutes each) while wearing 6 accelerometers: 2 each at the waist, thigh and shank. Accelerometer and video data were collected during activity.

Results: Excellent reliability and validity were found for measurements taken from accelerometers mounted at the waist and shank during running (Reliability: intraclass correlation (ICC) ≥ 0.99; standard error of measurement (SEM) ≤1.0 steps; Validity: Pearson ≥ 0.99) and at the thigh and shank during bicycling (Reliability: ICC ≥ 0.99; SEM ≤1.0 revolutions; Validity: Pearson ≥ 0.99). Excellent reliability was found between measurements taken at the waist and shank during running (ICC ≥ 0.98; SEM ≤1.6 steps) and between measurements taken at the thigh and shank during bicycling (ICC ≥ 0.99; SEM ≤1.0 revolutions).

Conclusions: These data suggest that the GT3X+ can be used for measuring step-count during running and pedal-revolution count during bicycling. Only shank placement is recommended for both activities.

Journal

Journal of Sports Sciences


Categories